IVF Laminar Airflow

  • Aeromech™ offers international standard IVF Version Vertical Laminar Flow Cabinets that have been designed specifically for polymerase chain reaction applications. Different than conventional dead air boxes, this enclosure features ISO Class 5 (US Federal Standard 209E Class 10) product protection standard within the work zone by vertical laminar airflow. The system is unique, superior and economical alternative to conventional dead-air boxes and passive chambers, which are traditionally used for IVF applications.
  • The IVF Version Vertical Laminar Flow Cabinets provide a particulate-free work environment and isolates critical steps in the IVF process from contamination present in the ambient environment. The system can be operated effectively as "mini-environments" in clean rooms, and also in laboratory. The system has a positively pressurized interior ensuring that contamination from the ambient environment cannot enter the work zone.
1 Product Aeromech® Vertical laminar Airflow – IVF Version
2 Application In- Vitro Fertilisation Application
3 Dimensions As per drawings enclosed.
4 MOC Galvanized Iron with PU coated / Stainless Steel – Satin Finish / Wooden laminated
5 Pre Filter Aeromech Make Pre Filter with Micro Glass Fibre Media of EU - 4 Rating.
6 Supply HEPA Filter Imported Minipleat HEPA Filter efficiency 99.97% Down to 0.3µ H-13 Rating.
7 Motor Blower Assembly Induction Motor with Centrifugal Blower Statically and Dynamically balanced with Suitable Rating and Size.
8 Mounting Unique Spring Suspension Systems to take care of Vibration and Noise.
9 Working Table S.S. 304 with Stabilised Heating Table with Microscope Arrange ment
10 Side Panel Accessory No. Clear Glass Side Panels/ Clear Acrylic panel/ UV Protected Polycarbonate panel
11 Pressure Gauge DWYER, USA Minihelic Differential Pressure Gauge/ Aeromech Inclined manometer.
12 Lights Flourescent Tube with Milky White Diffuser.
13 Other Electricals 3 Pin Socket, 15 AMP - 1 No. - On / Off Switches with Indicator.
14 Other Electricals 3 Pin Socket, 15 AMP - 1 No. - On / Off Switches with Indicator.
15 Air Cleanliness Class ISO - 5 (as per ISO - 14644-1).
16 Air Flow Rate 0.45 ± 0.05mps. or 90 FPM ± 10 FPM
17 Illumination Of Light Minimum 800 Lux on work table
18 Noise Level Less than 67 db on Scale 'A' When Ambient sound Level is not Greater than 57 db on Scale 'A'
19 Vibration Level Minimum.