Class II Biosafety cabinet – Type A2

The Aeromech Class II Biosafety Cabinet Type A2 is designed to protect the operator, product, and environment from pathogenic agents and airborne microorganisms, while also creating a sterile, dust-free air environment for safe handling of biological materials.
Class II Biosafety cabinet Type A2
High end Class II Biosafety cabinet Type A2 system provider

High-end Class II Biosafety cabinet – Type A2 system provider

Aeromech Class II Biosafety Cabinet Type A2 provides protection for the product, personnel, and environment.

These cabinets operate with a fan mounted at the top, which draws a curtain of sterile air over the workstation where biological products are handled.

The air then circulates underneath the workstation, returns to the top of the cabinet, and passes through HEPA filters.

The exhaust air is drawn in from the front of the cabinet and underneath the work surface, acting as a barrier to prevent potentially contaminated air from escaping toward the operator.

Versatile Class II Biosafety cabinet – Type A2 Supplier in India

We supply high-end Class II Biosafety cabinet – Type A2 for various industries, ensuring top-quality performance and service. Our advanced systems provide exceptional contamination control, enhancing the cleanliness of your critical environments.
Salient Features of Class II Biosafety cabinet Type A2

Salient Features of Class II Biosafety cabinet – Type A2

HEPA Filtration System

Removes 99.99% of particles, ensuring a sterile working environment

Operator and Product Protection

Ensures safety by protecting the operator, product, and environment from biological and chemical contaminants.

Airflow Dynamics

Balanced airflow prevents contamination while maintaining a sterile workspace.

Energy Efficiency

Energy-efficient components reduce costs without compromising performance.

Ergonomic Design

Offers a comfortable, user-friendly workspace with low noise, bright lighting, and easy accessibility.

Customized Class II Biosafety cabinet Type A2 Manufacturer

Customized Class II Biosafety cabinet – Type A2 Manufacturer

Aeromech Class II Biosafety Cabinet Type A2 cabinets have a minimum inflow velocity of 100 ft/min. In these cabinets, air enters the chamber through the front aperture, protecting the operator.

The inflow air mixes with the downflow air from the top of the cabinet and enters the front intake grille. The air then passes over the workstation, where it splits: approximately 60% to 70% of the contaminated air is recycled back into the workstation through the downflow HEPA filter, while the remaining 30% to 40% is exhausted through the exhaust HEPA filter.

Product Technicalities

Explore technical specifications and features of Class II Biosafety cabinet – Type A2, ensuring optimal performance and reliability for effective contamination control in critical environments.
Product:

Class II Biosafety Cabinet –Type A2

Air balancing:

70% recirculation.30% exhaust

Protection:

Product & Personnel & Environmental protection

Bio safety Level:

Bio safety Level 1 , 2 & 3

Exhaust Requirement:

Return to Laboratory or Thimble Ducting Only

Cleanliness level:

ISO class 5 as per ISO 14644-1

Average Airflow:

Inflow Velocity Minimum 0.45 m/s ( 90 FPM) at 8” front opening

Velocity:

Down Velocity 0.30 m/s ( 65 FPM)

Noise level:

60 db on A scale

Light Intensity:

Minimum 900 lux inside the working area at ambient zero

MOC:

Main Body –  Galvanized Iron with PU coated / Stainless Steel – Satin Finish
Working Area –  Stainless Steel IS 304 Grade Satin Finish

Supply Filter:

HEPA Filter having EU 13 rating with suitable size.

Efficiency:

99.99% down to 0.3 Microns

Exhaust Filter:

HEPA Filter having EU 13 rating with suitable size.

Efficiency:

99.99% down to 0.3 Microns

Pre Filter:

Aeromech® make, EU 4 rating with suitable size.

Efficiency:

90% down to 10 Microns.

Noise Level:

Less than 70 db ‘A’.

Vibration Level:

Minimum

Magnehelic Differential Pressure Gauge:

01 No. (make : Dwyer, USA) with each unit 
(To measure differential pressure across the supply HEPA filter).
Statically & Dynamically balanced with suitable rating and size.

Motor Blower Assembly:

(Induction motor with centrifugal blower) A Mounting unique spring suspension system to take care of vibration and noise.

3 Pins Switch & Socket:

15 AMPS – 1 set in each unit

Other Switches:

On/Off switches for motor blower & CFL lights

Explore the comprehensive specifics of Class II Biosafety cabinet – Type A2
Top Quality Clean Room Solutions from Aeromech Equipments

Best Class II Biosafety cabinet Type A2 Manufacturers in India

As the leading cleanroom engineers in India, we deliver premium Class II Biosafety cabinet Type A2, ensuring maximum effectiveness for superior contamination control and operational excellence.

We are the superior Class II Biosafety Cabinet Manufacturers in Delhi, Bengaluru, Hyderabad, Mumbai, Kolkata, Chennai, Trivandrum, Pune, Bhubaneswar, Chandigarh, Madurai, Visakhapatnam, Coimbatore, and Cochin.

Why Choose Aeromech?

We deliver exceptional quality and reliability in cleanroom solutions, backed by extensive experience and industry-leading expertise.
NABL Accreditation

ISO/IEC 17025:2017 testing laboratory certified

Cleanroom Validation

Validated over 600,000+ SQ FT areas.

Equipment Installations

Installed 3000+ cleanroom systems all over India.

Biosafety Cabinet Certification

Accredited Class II biosafety cabinet certifier.

Cleanroom Testing

CTBI-I cleanroom testing accreditation awarded.

Biosafety Cabinet Certifications